(Fixed PPM) Model 7100 – 0 – 2,000 Parts Per Million (PPM) Range Oxygen Analyzer
The Ntron Model 7100 provides an affordable and versatile Oxygen Analyzer solution that’s ideal for high-purity gas manufacturing, high quality welding applications, semiconductor manufacturing applications and any other application that requires precise measurement of oxygen. The unit provides high-purity gas measurement in a range of 0 – 2,000 ppm.
A small footprint and compact size makes the Ntron 7100 easy to install. The unit’s expected operational life is five to ten years. The low-maintenance, long-life Zirconia sensor has virtually unlimited shelf life and will not dry out or freeze.
Specifications
Sensor: | Mini Zirconia Sensor |
Measurement Range: | 0-10PPM/0-100PPM/0-1,000PPM/0-2,000PPM/auto range or fixed range |
Display: | 0.75” (19.05mm) LED digital display, Color coded LEDs for system status, Alarms 1&2= RED, system fault= YELLOW, OK and online= GREEN |
Response Time: | <15 sec. to 90% step change @ PPM level |
Intrinsic Error (accuracy): | ±3PPM @ 0-50PPM/±5PPM @ 51-500PPM/±5PPM @ 501-2000 PPM O2 @ STP when calibrated @ target concentration @ STP: ±1% range |
Operating Temperature: | 5-45º C |
Signal Interface: | RS-232 service port, 4-20mA negative ground, 0-10VDC range ID and choice of 0-1VDC or 0-5VDC or 0-10VDC analog oxygen signal |
Alarm Outputs: | Two alarms with adjustable form C relay outputs. Configurable for fail-safe or fail-alarm mode, ascending or descending trip. One system fault relay. One alarm/relay for sensor heater OK. |
Sample Port: | 1/8” NPT female ports |
Electrical Connections: | Removable terminal blocks on back of analyzer |
Power, Analyzer: | 5W at 115/230VAC 50/60Hz, single phase or 12-30VDC (battery backup capability) |
Environmental Ratings, Analyzer: | Faceplate: NEMA type 4, IP66 Housing: NEMA type 1, IP20 Warranty: Electronics and sensor: One yr. |
Weight: | Analyzer: Less than 2 lb. (0.9 kg) |